2. Safety

Knife guard on the knife holder
Each knife holder is equipped with a tightly

mounted knife guard (8, 9). This makes it possible

to cover completely the cutting edge in every

knife or blade position.

Knife holder N/NZ
The knife guard (8) of the knife holder N/NZ

can be easily positioned via the two handles (7)

(

Fig. 5

).

To cover the knife edge, push both cover strips

of the knife guard to the center.

Knife holder E
The knife guard on knife holder E consists of a

red foldaway handle. To cover the cutting edge,

fold the knife guard handle (9) upwards as illus-

trated in

Fig. 6

.

7

7

8

Centering the handle
For safety reasons, the handle of the handwheel

must always be centered while in motorized

sectioning mode.

• Activate the handwheel lock.

• To center the handwheel, lightly pull the han-

dle (6) outwards and swivel it into the center

of the handwheel (12) (

Fig. 4

).

• The handle will then lock automatically when

released.

9

10

11

6

12

12

Fig. 4

Fig. 5

Fig. 6

For the new knife holder E, the two

clamping levers (10, 11) must always

remain in the position shown. Clamp-

ing lever for the blade (10) at the right,

clamping lever for the lateral displace-

ment (11) at the left.