2. Attach vertical column

• Insert the cap screw (b) into the hole of the

baseplate from below. Place the lock washer

(c) on the cap screw from above.

• Thread the silver vertical column (2) onto the

baseplate from above and tighten with the

Allen key size 8.

A

3. Attach horizontal arm

• Slide the support ring (e) onto the vertical

column and position it so that the lock nut (f)

is facing to the back. Tighten the lock nut.

• Slide the cross-member (d) onto the column.

Ensure that the lock screw (g) is facing to the

right of the baseplate. The horizontal arm must

be centered over the microscope.

• Slide the horizontal arm (3), flat side facing the

lock screw (g), into the cross-member (d) and

tighten.

Fig. 63

Fig. 62

Fig. 61

b

c

2

e

d

2

g

f

h

For more information on connecting and

using the microscope, magnifying lens

or cold-light source, please see the ap-

propriate Instructions for Use.

Extreme danger of tipping! After install-

ing the vertical column, immediately

place the microtome onto the baseplate

so that the front feet of the microtome

are positioned in the shallow recesses

(A).

The lock washer must be positioned

between the baseplate and vertical col-

umn to prevent unintentional turning of

the column.