7 tray, 8 freezer pack, Freezer pack – Leica Biosystems RM2255 User Manual

Page 62: Optional accessories

Advertising
background image

60

Instructions for Use, V 1.8 RevC – 07/2013

The freezer pack consists of the freezer plate (98)

and the insulation jacket (99). It is for cooling (or

keeping cold) paraffin specimens. Up to 48 stan-

dard cassettes can be cooled at the same time.

The specimens require approx. 20 min. for being

cooled from room temperature (approx. 20 °C) to

"sectioning temperature."

The cooling performance depends on the ambi-

ent temperature and the volume of the speci-

mens being cooled.

The insulation jacket can be placed on the hood

of the microtome in the same manner as the tray,

or it can be placed next to the instrument.

6.8 Freezer pack

98

99

98

The tray is mounted on the hood of the micro-

tome so that the small feet on the underside fit

into the cutouts on the hood.

It is for storage of the utensils used during sec-

tioning as well as the sectioned specimens.

6.7 Tray

When using the freezer pack on the in-

strument the tray (Fig.

47

) must be re-

moved.

• Remove the freezer plate from the insulation

jacket and place it in a deep freezer, ideally

overnight (but for at least six hours) at approx.

–23 °C.

• Reassemble the freezer plate and insulation

jacket and place them on the bench or the

instrument.

Fig. 56

Fig. 57

6.

Optional Accessories

Advertising