5.10 Sectioning

• Center the handle of the handwheel.
• Use the TRIM/SECT key to select the sectioning mode.
• Adjust the appropriate sectioning thickness or verify the selected value.
• Depending on the operation planned, use the CUT MODE key to select

one of the motorized sectioning modes, CONT, SINGLE or STEP.

+

5.11 Changing the specimen or interrupting sectioning

Lock the handwheel and cover the knife edge with the knife guard prior to any manipula-

tion of knife or specimen head, as well as prior to changing the specimen block and dur-

ing all work breaks!

• Raise the specimen to the upper end position and activate the mechani-

cal handwheel lock.

• Cover the sectioning edge with the knife guard.
• Remove the specimen from the specimen clamp and mount a new sam-

ple to continue.

• Before cutting into a new specimen, move the specimen head back to

the rear end position.

• Check the sectioning speed setting and set an appropriate speed.
• Start motorized sectioning.
• Pick up the sections and mount them on microscope slides.

Always set the sectioning speed according to the hardness of the

specimen! For hard specimens, always select a slow speed.

Always use a different area of the cutting edge for trimming and sectioning. To do so, laterally

displace the blade or knife in the knife holder.

When using the knife holder E with lateral displacement, it is sufficient to move the knife