Operation – Leica Biosystems RM2255 User Manual

Page 34

Advertising
background image

32

Instructions for Use, V 1.8 RevC – 07/2013

• Using the CUT MODE button, select the desired operating mode, CONT,

SINGLE or STEP, on the control panel (Fig. 18).

CONT (continuous stroke) operating mode
• Press the foot switch once briefly to start motorized sectioning.

5. Operation

• Press the foot switch again to stop it.

The specimen stops in the next upper end position.
SINGLE (single stroke) operating mode
• Press the foot switch once briefly to start motorized sectioning. After

every step, the specimen stops automatically in the upper end position.

STEP (step stroke) operating mode
• Press down the foot switch to start the sectioning process. The speci-

men is now moved for as long as the foot switch is depressed.

• If the foot switch is released, the specimen remains stationary in the

position that it has reached.

How to activate the emergency stop function
• Press the foot switch strongly to activate the emergency stop function.

Sectioning stops immediately.

The red LED in the E-STOP field on the instrument (Fig. 14) is lit up as long

as the foot switch remains depressed.

• To continue, restart the sectioning process using the foot switch. The

operating mode remains unchanged.

If the foot switch remains pressed for longer than half a second, the

specimen stops in the next upper end position.

Caution!

In addition to the foot switch, all control panel functions and all but-

tons on the instrument continue to be active.

Foot switch (optional accessory)

The foot switch can be used to control the motorized sectioning process. It

also has a function that is similar to the emergency stop function.