25

Leica RM2255

When continuous feed is selected, the coarse feed buttons have the same functions

as in trimming mode. The STEP function is useful for careful step-by-step approxi-
mation of the specimen towards the blade.
How to activate the STEP function:
• Switch the instrument on while holding the

button on the control panel.

(Likewise, to deactivate switch the instrument on while pressing the

button.)

While the instrument is initializing, hold the

button until the software version

number is no longer displayed (ensure that version number 2.1 was displayed).

• Press the TRIM/SECT

button and select sectioning mode (LED SECT lit).

• When pressing the button (with an arrow) for slow-speed coarse feed, a

defined, incremental feed takes place (STEP) at the value indicated on the

display in the respective direction (simple step).

• By short activation of the buttons for coarse feed (with two arrows), a single

step is also effected in the appropriate direction.

• Longer activation of the double-arrow coarse feed button effects a repeated

feed motion for as long as the button is pressed.

5. Operation

To start the rapid backwards movement (away from the blade), press the

key.

After the

button is pressed, the specimen head is moved to the rear end

position.

• To stop the movement, press any of the four coarse feed buttons.

• The yellow LED (40) in the button flashes while the specimen head is in mo-

tion, and remains lit continuously when the rear end position is reached

.

40

Backwards coarse feed

Multiple

step

backward

Multiple

step

forward

Single

step

backward

Single step

forward

Sectioning mode

Trimming mode

Button functions in STEP

mode

Do not put your fingers between the specimen clamp and micro-

tome to prevent pinching them.

In the trimming mode, the coarse feed buttons operate a continuous

movement as long as the button is held down. The double-arrow

button for rapid coarse feed backward movements has a lock-in

function.

This feature is only available in sectioning mode in version 2.1 or

higher. For software versions older than 2.1, contact Leica Techni-

cal Service.

In sectioning mode the user can select between STEP function

(step-by-step specimen feed) and continuous specimen feed.