31

Leica RM2255

5. Operation

Indication of remaining horizontal feed

The visible and audible remaining feed indication

feature informs the user during trimming and

sectioning when a remaining feed of approxi-

mately 1 mm is available before the front limit is

reached.
The yellow LED (41) in the COARSE FEED button

lights up from the beginning of the remaining

feed. In addition, an acoustic signal is heard for

approx. 2 seconds.
The sectioning process is interrupted and the

object head stops in the upper end position.
From this point on, a remaining feed of approx.

1 mm is available.
In the remaining feed area, no more object feed-

ing to the knife is possible using the coarse feed

buttons.

• Restart motorized sectioning.

The yellow LED (41) in the COARSE FEED button lights up.
• When the front end position is reached, the sectioning process stops

automatically.

• Upon restart, no more feed motion takes place.
• You can continue to work on the specimen by pressing the correspond-

ing coarse feed button in the rear end position (HOME) and continuing

with sectioning

.

Specimen head

and

HOME

41

or

Remaining travel

Total feed

To do so, you must press TRIM/SECT to switch to trimming mode, as

otherwise, you cannot use the coarse feed.

If the specimen head is already in the remaining feed range when

the instrument is switched on, an additional acoustic signal is

heard after the software version is displayed.

• You can continue to work on the specimen by moving it back a short

distance using the coarse feed buttons (set trimming mode!).

• The STEP function is disabled in the remaining feed range.

Fig. 19