2 instrument specifications, Instrument specifications, Instrument components and specifications – Leica Biosystems RM2255 User Manual

Page 14

Advertising
background image

12

Instructions for Use, V 1.8 RevC – 07/2013

3.

Instrument components and specifications

Fig. 8

Motor

protection

fuse

Heat sink

Voltage selector switch

with fuses

Main switch

Power supply

Connecting

cable for control

panel

Connecting jack

for foot switch

3.2 Instrument specifications

The Leica RM2255 is a motorized rotary micro-

tome.
• The specimen feed system with zero-backlash

and maintenance-free cross roller guides and

the stepper motor operated coarse feed sys-

tem are located in a dust-proof plastic housing.

• The instrument is equipped with a safety hand-

wheel with a handle that can be centered, as

well as a mechanical handwheel lock.

• All controls and LEDs are centralized in a

separate control panel.

All control elements are logically arranged in

functional groups and easily identifiable.

• The specimen retraction can be turned off.

In manual operation the retraction can be

adjusted. In motorized operation the retrac-

tion value varies with the selected sectioning

speed. An LED illuminates while the sample is

in retraction.

• The electric coarse feed operates at two

speeds. In the sectioning mode, the coarse

feed buttons have a STEP function.

• Three motorized sectioning modes (CONT,

SINGLE, and STEP) are available as well as

one manual sectioning mode, the rocking

mode ROCK.

In rocking mode, it is enough to move the

handwheel a short distance back and forth in

order to create a section.