7. Troubleshooting

Problem

Possible cause

Corrective action

• The blade is dull.

• The pressure plate is damaged

or incorrectly adjusted.

• Clearance angle of the knife/

blade too small.

• Laterally displace the knife

holder or insert a new blade.

• Insert a new pressure plate or

use a new knife holder.

• Methodically experiment with

larger clearance angle settings

until you have found the optimum

angle.

2. Section compression
The sections are very com-

pressed, show folds or are

squeezed together.

• The blade is dull.

• The specimen is too warm.

• The sectioning speed is too fast.

• Use another area of the blade or

a new blade.

• Cool the specimen before sec-

tioning.

• Reduce the sectioning speed.

3. "Stripes" in sections
For knife holder E

• There is an accumulation of par-

affin on the rear pressure plate

of the knife holder.

• Remove paraffin from this area

on a regular basis.

4. Noises during sectioning
The knife "sings" when section-

ing hard specimens. The sec-

tions show scratches or chatter

marks.

• The sectioning speed is too fast.

• The clearance angle is too

wide.

• Insufficient clamping of speci-

men clamp and/or knife holder.

• Select a slower sectioning

speed.

• Methodically reduce the clear-

ance angle setting until you have

found the optimum angle.

• Check all screw and clamp con-

nections on the specimen holder

system and the knife holder. If

necessary, tighten the levers and

screws.

5. High blade consumption

• Too great of a sectioning force

was applied.

• Adjust the sectioning speed and/

or section thickness when sec-

tioning. Select a thinner section.