5. Operation

Configuring the retraction settings

• To call up the retraction settings, press the MENU MODE and CLEAR

buttons simultaneously.

• The current set value is displayed as a three-digit number in the four-

digit display (e.g. "025" = 25 µm).

• Select the desired retraction value.

The retraction value can be adjusted in increments of 5 µm to a maximum

of 100 µm using the

-

buttons on the control panel.

• To exit the retraction settings, press MENU MODE.

A retraction movement takes place by the newly selected value after

each section.

• While the specimen is in retraction, the yellow LED on the RETRACT

display (4) lights up.

Switching off the specimen retraction

• To call up the retraction settings, press MENU MODE and CLEAR simul-

taneously.

• To switch off retraction, press the button on the control panel until

the display indicates "OFF".

• To exit the retraction settings, press MENU MODE.

When retraction is switched off, the specimen is not retracted.

The yellow LED (4) of the RETRACT indicator does not light up.

+