Instrument components and specifications – Leica Biosystems RM2255 User Manual

Page 16

Advertising
background image

14

Instructions for Use, V 1.8 RevC – 07/2013

3.

Instrument components and specifications

Microtome

Sectioning thickness setting:
Section thickness setting range:

Setting values:

0.50 - 100 µm

from 0.50 - 5.0 µm

from 5.0 - 20.0 µm

from 20.0 - 60.0 µm

from 60.0 - 100.0 µm

in 0.5 µm increments

in 1.0 µm increments

in 5.0 µm increments

in 10.0 µm increments

Trimming section thickness setting range:

Setting values:

1 - 600 µm

from 1.0 - 10.0 µm

from 10.0 - 20.0 µm

from 20.0 - 50.0 µm

from 50.0 - 100.0 µm

from 100.0 - 600.0 µm

in 1.0 µm increments,

in 2.0 µm increments,

in 5.0 µm increments,

in 10.0 µm increments,

in 50.0 µm increments.

Specimen feed:

28 mm ±1 mm, feed motion via step motor

Vertical stroke:

70 mm

Maximum sectioning area w/o retraction:

65 mm without specimen orientation

Maximum sectioning area with retraction:

60 mm

Specimen retraction:
in manual sectioning mode:

5 - 100 µm in 5 µm increments; can be turned off

in motorized sectioning mode:

Varies with the sectioning speed; can be turned off

Electric coarse feed:

300 µm/s and 900 µm/s

Sectioning speed:

0; 0.5 - 420 mm/s ± 10 %

Return stroke speed:

approx. 120 - 420 mm/s ± 10 %

Repositioning of knife holder base

North-south:

East-west movement:

± 24 mm

± 23 mm

Maximum specimen size (L x H x W):

50 x 60 x 40 mm

Specimen orientation

Horizontal:

Vertical: