Operation – Leica Biosystems RM2255 User Manual

Page 35

Advertising
background image

33

Leica RM2255

Directional specimen holder fixture

Orienting the specimen

Display of the zero position
For better display of the zero position, the orien-

tation has two red indicators (32).
When both indicators are visible and both set-

screws (30, 31) are in zero position at the same

time (click stop, white marking on ""), the speci-

men is in zero position.

5. Operation

The directional specimen holder fixture may be

exchanged for a non-directional fixture (optional

accessory).

The specimen orientation allows for simple posi-

tion correction of the specimen surface when the

specimen is clamped into place.

32

30

32

31

29

Fig. 20

When the large standard specimen

clamp (50x55 mm) is used, the speci-

men orientation of 8° in north-south

direction is no longer possible.

The usable angle is only about 4° in this

case.

In the quick clamping device of the directional specimen holder fixture, all specimen

clamps available as optional accessories can be used (implemented).

• Raise the specimen head to the upper end

position and activate the handwheel lock.

• To release the clamp, turn the eccentric lever

(29) forwards.

• Turn setscrew (30) to orient the specimen in

north-south direction. Turn setscrew (31) to

orient the specimen in east-west direction.

Each complete turn of the screw inclines the

specimen by 2°. A total of 4 complete turns = 8°

are possible in every direction. The accuracy

is approximately ± 0.5°.

For ease of estimation, there is a white mark on

the handle and a click stop that is noticeable

during turning.

• To lock the current orientation, turn the

eccentric lever (29) backwards.

Specimen blocks must NOT be oriented

during the retraction phase!

If a block is oriented during retraction,

the block will advance by the retraction

value PLUS the selected section thick-

ness before the next section.

This may cause damage to both speci-

men and knife.

Advertising