52

Instructions for Use, V 1.8 RevC – 07/2013

Inserting the blade, knife holder E and E-TC
• Fold knife guard (9) downward.
• To insert the blade, flip the right clamping lever

(10) forward and down.

• Carefully insert the blade from the side. Make

sure that the blade is clamped parallel to the

upper edge of the pressure plate.

• To clamp the blade, rotate clamping lever (10)

back upwards.

6.3.2 Knife holder E/E-TC

9

11

10

Knife holder E-TC

Knife holder E

9

10

9a

The knife holder E-TC is designed

for the Leica TC-65 tungsten carbide

blades.

Prior to inserting the blade, both knife

holder and knife holder base must have

been installed on the instrument!

Fig. 42

The knife holder E is designed for conventional disposable blades from all current manufactur-

ers. It is available in two models: one for low-profile blades and one for high-profile blades.

The knife holder has a lateral displacement, so that the entire width of the blade can be used.

The ejector (9a) ensures safe blade changing.

The clamping levers on the knife hold-

er are not interchangeable. The two

clamping levers (10, 11) must remain

in the position shown at all times, as

otherwise isolated malfunctions of the

knife holder can occur.

Clamping lever for the blade (10) at the

right, clamping lever for the lateral dis-

placement (11) at the left.

Fig. 43