Instructions for Use, V 1.8 RevC – 07/2013

Sectioning window setting

The sectioning window function allows to optimally adapt the size of the

sectioning area to the actual size of the specimen.

The selected sectioning window setting is maintained when the instrument

is turned off.
• Turn the handwheel to position the lower edge of the sample approxi-

mately 3 mm above the cutting edge.

• Press the "SET SECTIONING WINDOW" button. This defines the first

window border.

• The green LED (43) in the button flashes after the first window border is

defined.

• Pass the specimen through to the upper border of the cutting edge and

press the "SET SECTIONING WINDOW" button again.

• After the second window border is defined, the green LED in the button

goes out. This indicates that both values were accepted.

5. Operation

43

To set a sectioning window, you must always enter a pair of values.

The sequence of entry (upper or lower) is optional.

Canceling a defined sectioning window

• To cancel a defined sectioning window before the start, press the "SET

SECTIONING WINDOW" button once. This sets the sectioning area to the

maximum size (corresponding to the entire sectioning area).

The set sectioning speed is only active within the boundaries of the

sectioning area. A higher speed is used outside the sectioning area.

Setting the sectioning speed

• The sectioning speed can be set continuously (in the range of

0 - 420 mm/s) using the rotary knob.

The speed selector has a scale graduation from 1 to 10. The graduation is

provided for reference purposes and does not indicate a particular speed.