Troubleshooting, 1 instrument malfunctions, 1 error reports – Leica Biosystems RM2255 User Manual

Page 69: Instrument malfunctions

Advertising
background image

67

Leica RM2255

7.1.1 Error reports

If the message "E 1" follows the software version indication after switching

on the instrument, a loss of parameters has occurred.
• Press any key.
All parameters are then set back to the default values.
• Now all settings must be checked to see if the values are still correct.

The values may have to be reentered.
You can continue with your normal working routine.

7. Troubleshooting

7.1 Instrument malfunctions

7.1.2 Malfunctions, possible causes and troubleshooting

Certain instrument malfunctions appear on the 3-digit LED display or are indicated

through illuminated LED's on the control panel.

The following is a list of the most common problems which can arise while working

with the instrument, along with possible causes and troubleshooting procedures.

If the malfunction cannot be remedied with any of the options listed in the table, or the

problem occurs repeatedly, inform Leica Service Support immediately.

Problem

Possible cause

Corrective action

• No display; no reaction to but-

tons pressed after the instru-

ment is switched on.

1. The power cable is not con-

nected correctly.

2. Power fuses are defective.

3. The cable of the control panel

is not connected correctly.

4. The voltage selection switch is

set incorrectly.

1. Check the connection of the

power cable.

2. Replace the power fuses.
(see

Chapter 8.2.1

)

3. Check the connection of the

cable to the control panel.

4. Check the voltage setting and

correct it if necessary

(see

Chapter 4.5

)

Advertising