13

Leica RM2255

3.3 Technical Data

3. Instrument Components and Specifications

General

Approvals:

The instrument-specific marks are located next to the identi-

fication label.

Nominal supply voltages:

100 / 120 / 230 / 240 V AC ±10 %

Nominal frequency:

50/60 Hz

Maximum power consumption

340 VA

Protection class

:

I -

according to IEC-1010, UL 3101, EN 61010

Power fuses

2 x T 3.2 A, UL listed

Pollution degree

:

2 -

according to IEC-1010, UL 3101, EN 61010

Overvoltage category

II -

according to IEC-1010, UL 3101, EN 61010

Maximum heat emission:

340 J/s

Operating temperature range:

+10 °C to +35 °C

Temperature range during storage:

+5 °C to +55 °C

Relative humidity:

max. 80 %, non-condensing

Storage humidity:

< 80 %

Dimensions and weight

Basic instrument

Width (including handwheel):

413 mm

Width (excluding handwheel):

300 mm

Depth (including waste tray):

618 mm

Height (total):

305 mm (with tray on the hood)

Working height (knife blade):

100 mm (measured from the base plate)

Working height (knife blade):

168 mm (measured from the table)

Weight (without accessories):

approx. 37 kg

Control panel

Width:

121 mm

Depth:

166 mm

Height:

50 mm

Height (in inclined position):

81 mm

Weight (net):

approx. 0.660 kg