Troubleshooting – Leica Biosystems RM2255 User Manual

Page 70

Advertising
background image

68

Instructions for Use, V 1.8 RevC – 07/2013

28

Problem

Possible cause

Corrective action

• The cutting motor cannot be

started.

1. The EMERGENCY STOP button

is pressed or the foot switch is

depressed.

2. The handwheel lock is acti-

vated.

3. The motor was temporarily

overloaded and was switched

off via the automatic fuse.

4. The foot switch or dummy plug

is not connected or not con-

nected correctly.

1. Unlock the EMERGENCY STOP

button and select the operating

mode; release the foot switch.

2. Release the handwheel lock.

3. Switch off the instrument and

wait 30 seconds.

Push the button of the auto-

matic fuse (28) on the rear of

the instrument.

Switch the instrument back on.

4. Connect the foot switch or

dummy plug and ensure that it

is connected correctly.

• Feed motion no longer occurs.

The cutting motor cannot be

started.

1. The end of the residual path

has been reached.

2. The specimen was already

within the residual path when

the instrument was switched

on.

1. Select the trim mode by press-

ing TRIM/SECT. Then retract

the specimen using the coarse

feed button.

2. Follow the procedure in Item 1.

• The coarse feed buttons can

be used to move the object in a

backwards direction only.

1. The motorized sectioning pro-

cess is still in progress.

1. End motorized sectioning by

pressing the foot switch or the

RUN/STOP or ENABLE button.

• The specimen in the residual

path area cannot be returned

using the coarse feed buttons.

1. The STEP mode is still active

(LED in SECT button lights up).

1. Press the TRIM/SECT button

to switch to trimming mode.

Return the specimen using the

coarse feed button.

• Cutting motor switches off

prematurely in CONT (continu-

ous stroke) operating mode.

1. Foot switch was held down too

long when starting the section-

ing process.

2. The RUN/STOP and ENABLE

buttons of the control panel

were not released simultane-

ously after starting.

1. Only press the foot switch

briefly to start the sectioning

process.

2. Release both buttons simulta-

neously.

• HOME and STOP are lit up

simultaneously.

1. The end position detection is

defective.

1. Contact Leica after-sales ser-

vice.

7. Troubleshooting