Applying actuator power – MTS Acumen User Manual
Page 189

Interlock Info Tip on System Panel
Caution
:
Clearing interlocks overrides all interlocks and may result in unexpected actuator motion.
Unexpected actuator motion may result injury or damage to equipment.
Always make sure the actuator area is clear and double check your work before clearing
interlocks.
Warning
:
Clicking the Override Interlock button will override ALL limits for 30 seconds each time the
button is pressed including the Axial Acceleration Limit which is designed to limit damage due
to unexpected actuator motion.
Unexpected actuator motion may result in injury or equipment damage.
Use of the Override Interlock function is not recommended for Acumen systems.
Applying Actuator Power
Warning
:
Applying power can result in sudden actuator movement.
A moving actuator can damage objects or injure anyone in its path.
Always clear the actuator area before applying power.
Use the Actuator Power panel to provide power to the actuator. You must apply low power before you can
apply high power. When an interlock is triggered, the power buttons are disabled until you reset the interlock.
If a controller is not connected or frame power is off, the actuator power buttons are disabled.
MTS Acumen
™
| 189