Frame-mounted control – MTS Acumen User Manual

Page 38

Advertising
background image

Description

Item

Interlocked.

Red (Solid)

System is in service mode. To move the system out of
service mode, slide the Service Mode switch on the upper

Red or White (Blinking)

back of the system, and click Interlock Reset. If problems
continue, contact MTS Technical Support.

Interlock cleared, standby power is on.

White (Solid)

Interlock cleared, low power.

Blue (Blinking)

Interlock cleared, high power, not running.

Blue (Solid)

Interlock cleared, running state.

Green (Solid)

Interlock cleared, temperature warning.

Yellow (Solid)

Note:

The fans to cool the actuator increase in speed with
increasing temperature and will become louder. This
is normal system operation, and it is not necessary
to shut down the system. Reducing the frequency
and/or force output of the system will allow the
actuator to cool down. The system status light will
remain yellow until the actuator has cooled down.

Frame-Mounted Control

The frame-mounted control can be attached to the left or right side of the frame. It provides controls to
help you mount fixtures and install specimens. The frame-mounted control also has an alphanumeric
display and illuminated icons to provide feedback.

Note:

When exclusive control is provided by the frame-mounted control, the actuator controls on the MP
application will be locked and overlaid by the frame-mounted control exclusive control icon:

The following figure separates the frame-mounted control into boxed sections that are described in the
table that follows.

38 | MTS Acumen

System Introduction

Advertising