Status panel – MTS Acumen User Manual
Page 49

Description
Icon
Number
Reset Interlock. Click to reset the interlock, or click the drop-down arrow
to select Reset Interlock or Override Interlock.
5
Override Interlock. Selecting this option overrides interlocks for 30
seconds. Use of this button is not recommended for Acumen systems.
Warning
:
Clicking the Override Interlock button will override ALL limits
for 30 seconds each time the button is pressed including the
Axial Acceleration Limit which is designed to limit damage
due to unexpected actuator motion.
Unexpected actuator motion may result in injury or equipment
damage.
Use of the Override Interlock function is not recommended
for Acumen systems.
To cancel an override, click the Reset Interlock button.
Frame status indicator. Blinks to indicate which load frame is associated
with the station when you select Controller > Identify Frame. This status
6
light corresponds to the Frame-Mounted System Status on the front hood
of the load frame. For more information about what each status light
indicates, see
(p. 36).
Click to open the Status Indicator window and view a list of status controls.
Red lights indicate which status indicators are triggered. For more
7
information about each status indicator and what you can do to resolve
triggered status indicators, see
(p. 49).
Status Panel
The Status panel shows a red indicator when a fault status has been triggered. Hover your cursor over
the red indicator on the panel to see a message explaining the error. Click the open window icon to see
a list of all status indicators. The following table describes the status indicators and the corrective action
to take when an error occurs.
MTS Acumen
™
| 49