Note: If your desktop does not have an MPX icon that designates the station configuration (for
example FT40_station1.cfg), see

Creating a Desktop Shortcut

(p. 144) for details regarding how

to set one up.

Position the Crosshead and Actuator for Initial
Setup

In this section, you will raise the crosshead high enough so that the actuator cannot contact anything even
if it goes to full extension. Next, you will retract the actuator to ensure that you have the correct motion
polarity and so that you know where the actuator end point is. Finally, you will position the actuator for the
type of test that you are doing.

1. Raise the crosshead high enough so that the actuator cannot contact anything even if it goes to full

extension.

Crosshead Locks and Lift Controls

Description

Item

Crosshead Locks (Locked)

1

Crosshead Lift Controls

2

Exclusive Control

3

98 | MTS Acumen

Running the Example Spring Test