Fine-tuning meter pressurization, Alternate skip/double check, Use the “ alternate skip/double check ” belo – Great Plains YP825A3P Operator Manual User Manual

Page 48

Advertising
background image

44

YP4-6-825A3P

Table of Contents

Index

Great Plains Manufacturing, Inc.

401-652M

Table of Contents

Index

2013-05-28

10. Resume planting.

A small varying population deviation between rows is
normal. If a row is consistently running lower or
higher than the other rows, it could indicate a meter
or seed sensor problem that needs attention.
Population Troubleshooting Charts” on page 67.

11. Continue to monitor meter pressurization.

When oil reaches operating temperature, fan speed
can change. Expect to make periodic adjustments to
fan circuit or butterfly valve to maintain ideal meter
pressurization.

12. Skips:

If all else is correct, and the overall average
population is running low (or there are gaps upon
furrow check), the cause may be seeds falling out of
disk pockets before delivery to the seed tube.
Increase meter pressurization to correct this.

13. Doubles:

If all else is correct, and the overall average
population is running high (or there are double seeds
upon furrow check), the cause may be two seeds in
some disk pockets at delivery to the seed tube.
Decrease meter pressurization to correct this.

Fine-Tuning Meter Pressurization

After several passes, and you are comfortable with the
planter’s operation, you can optimize meter
pressurization. Use this method during longer passes, so
there are fewer distractions (and see sidebar regarding
Milo).

14. Observe the current overall average population

reported by the monitor.

15. Adjust the fan speed down, in small steps, waiting

5-10 seconds between adjustments, until skips occur
(actual population begins falling below target). Note
the pressure at which skips begin.

16. Restore pressure to the initial value at step 14.

17. Adjust the fan speed up, by periodic small

increments, until doubles occur (actual population
begins rising above target). Note the pressure at
which doubles begin.

18. Adjust meter pressurization to a value halfway

between the limits established at step 15 and
step 17.
This is the ideal value for your crop, disk and
population, providing equal margin against skips and
doubles. Record this value for future use.

Insufficient meter pressurization, or unusually rough
fields, can increase the incidence of empty pockets. Be
sure to rule out other causes (such as skipping chains,
meter starvation, incorrect meter disks) before adjusting
meter pressurization to reduce apparent skips.

Excess meter pressurization can increase the incidence
of doubles. Be sure to rule out other causes (such as
incorrect meter disks) before adjusting meter
pressurization to reduce doubles.

Alternate Skip/Double Check
Small seeds, particularly Milo, are less reliably sensed as
skips or doubles by the seed tube sensors, and watching
monitor population is unlikely to locate the pressure
limits. This test can also be quicker for all seeds.

A. After planting a few passes with initial settings,

remove the rain covers from several rows (use rows
with a variety of seed hose lengths and routes).

B. Make a meter pressurization adjustment. Resume

planting for a pass or less.

C. Stop planter motion but leave engine at field rpm and

fan running.

D. Inspect the seed disks closely. Look for empty seed

cells (skips) and cells with multiple seeds (doubles).

E. Repeat step B-to-step D until limits are established.

Record limits. Re-install rain covers. Plant with
median settings.

Advertising
This manual is related to the following products: