Bio-Rad PDQuest 2-D Analysis Software User Manual

Page 51

Advertising
background image

6.2

Common Re-order Parts

Description

Catalog number

ProteomeWorks Plus Gel Sheets (thin plastic disposable sheets)

165-7057

ProteomeWorks Plus Cutting Mat (thick plastic sheet - approx. 3–4 mm)

165-7058

ProteomeWorks Plus 1.5 mm nominal Cutting Tip

165-7052

ProteomeWorks Plus 1.5 mm Eject Pin

165-7054

ProteomeWorks Plus 1.0 mm nominal Cutting Tip

165-7051

ProteomeWorks Plus 1.0 mm Eject Pin

165-7053

ProteomeWorks Plus Cutting Tip Wrenches

165-7056

ProteomeWorks Plus Calibration Target Template

165-7059

ProteomeWorks Plus Camera Alignment Target Template

165-7060

Immun-Blot PVDF Membrane, 25 x 28 cm

162-0255

Coomassie Brilliant Blue R-250 Staining Solution, 1 L

161-0436

Section 7
Maintenance and Adjustments

7.1

General Maintenance

When the System is not in operation, cover the instrument with the dust cover provided.
This prevents dust from collecting on the instrument.

For general cleaning of the system, wipe with a damp cloth. Ethanol or mild detergents
may be used. Do not use any harsh solvents or cleaners.

Replacing cutting platform lights.

The cutting surface lights are located directly beneath the cutting surface and consist of four 24W
fluorescent tubes. Appendix B has the diagrams showing the location of the fluorescent bulbs
and the location of covers that need to be removed. Remove the lighting front cover panel. The
required fluorescent tube(s) are gently pulled out and then replaced by pushing the replacement
tubes into position. After the tube(s) are changed the front cover is replaced.

Fuses

The main power fuse is located on the system box interface plate, refer to Appendix C for exact
location. The fuse for operation is a 3.15A fuse. Two of these fuses have been supplied with the
System.

7.2

Cutting Tip Cleaning

The cutting tip is made of high quality 303/304 stainless steel. If properly maintained, the cutting
tip will retain its cutting ability through repeated use. Replacement is needed if the cutting tip has
been deformed or if the spot pick-up rate is unsatisfactory even after cleaning and cutting tip
height adjustment.

If the pick-up rate is unsatisfactory, there may be gel material building up inside the tip. Either
clean out the tip with water or alcohol and a wire, or replace the cutting tip with a new one. Daily
maintenance includes washing the cutting tip with water or alcohol. Take the tip off the instrument
and squirt water or alcohol through the tip. To remove the cutting tip, see the following section on

48