Bio-Rad SmartSpec Plus Spectrophotometer User Manual

Page 43

Advertising
background image

The display will show either

or

Press the right arrow key to return to the assay.

After the last replicate of the last standard is measured, SmartSpec Plus will construct
the standard curve.

C. Editing standard curve data. After the absorbance of the last replicate of the last

standard has been read, SmartSpec Plus will offer you a chance to edit the data set
before constructing the standard curve. You can re-read a standard or any replicate of a
standard group, or you can edit the concentration information for any standard or standard
replicate group.

If you do not want to edit the data, press Enter and skip to D below. If you need
to edit the data or to re-read one or more samples, press Select to toggle from No
to Yes and press Enter.

To edit a concentration, press 1.

Press Select until the standard you want to edit is displayed and then press Enter.

Use the number keys to input the correct concentration and press Enter.

Press the left arrow key to return to the assay and complete construction of the
standard curve. If you want to re-read a standard, press 2.

Use the number keys to input the number of the standard that you want to re-read,
press Enter, input the replicate number and press Enter again.

40

Do you wish to modify

the Standard Curve? No

1. Edit Concentrations

2. Reread a Std

<=Back

Select Standard to edit:

Standard #1: 123.4567

A750 = X.XXX

Std #1 Rep #2 >= Continue

A750 = X.XXX = - X.XXX

Std #1 Mean >= Continue

Change Conc Std #1 from

123.45678 to

1. Edit Concentrations

2. Reread a Std

<=Back

Enter Standard#

Enter Rep#

Do you want to reread

blanks each time? Yes