Bio-Rad SmartSpec Plus Spectrophotometer User Manual

Page 46

Advertising
background image

If you want to rescale the Y axis, press Enter. If you want to rescale the X axis, press
Select to toggle the display and press Enter.

When you rescale the X axis, you must input the new maximum X value and press
Enter. A new maximum Y value is automatically calculated and the graph is plotted.

When you rescale the Y axis, you must input the new maximum Y value and press
Enter. A new maximum X value is automatically calculated and the graph is plotted.

Quantitating UV-proteins:

Protein (mg/ml) = C1 * (A

280

– A

320

) – C2 * (A

260

-A

320

)

Coefficients 1 and 2 (defaults 1.55 and 0.76, respectively) can be changed for different pro-

teins.

43

Max conc=150.0000

Enter new max:

Max abs=1.234

Enter new max:

Rescale standard curve?

Absorbance (X-axis)