Trim leading edge of paper roll with scissors if uneven or torn.

Feed paper into slot opening until it stops. Press the "Paper Feed" key located on the
keypad and guide the paper into the printer until it grabs and self feeds.

Continue to press "Paper Feed" button until paper comes out of printer outlet located on
top of case as shown.

Wind any excess paper back on roll and drop new paper roll with shaft into grooves in
paper storage area as shown.

Re-install printer door cover.

Tearing Paper

When tearing off print outs, always tear paper forward against serrated edge of printer
outlet as shown on previous page.

Section 10
Specifications

Characteristic

SmartSpec Plus Specification

Light Source

Xenon lamp

Detector

Photodiode array

Display

2 x 24 alphanumeric backlit LCD

Printer

Built-in alphanumeric printer

Interface

RS-232 serial port

Wavelength Range

200 to 800 nm

Wavelength Accuracy (18–27°C)

< ±1.0 nm from 200–250, 300–800 nm
< ±0.5 nm from 250 to 300 nm

Spectral Bandwidth

<5 nm

Absorbtion Accuracy

±0.01 @ 0.5 AU
±0.02 @ 1.0 AU

Absorbtion Repeatability

±0.005 @ 0.5 AU

Stray Light

<0.2% @ 230 nm using NaI Standard

Operating Temperature

15 to 35°C

Beam Height (Z-dimension)

8.5 mm

Dimensions

33 x 26 x 14 cm (W x D x H)

Weight

~3.9 kg

Power Requirement

90–260 VAC, 47–63 Hz

Regulatory

CE, TЬV