11.3 Cuvette Selection Guide

Catalog

Minimum

Maximum

Cuvette

Number

Volume

Volume

Type

Pathlength

170-2502

1,000 µl

3,500 µl

Standard cuvette, quartz

10 mm

170-2503

500 µl

1,400 µl

Semi-microvolume cuvette, quartz

10 mm

170-2504

200 µl

700 µl

Microvolume cuvette, quartz

10 mm

170-2505

80 µl

100 µl

Sub-microvolume cuvette, quartz

10 mm

11.4 Selecting the Appropriate Cuvette

The SmartSpec Plus has 1 mm, 2 mm, 5 mm and 10 mm pathlengths. For most

absorbance measurements you will be using 10 mm pathlength cuvettes. Cuvettes of shorter
pathlength allow the measurement of concentrated samples without dilution. If you put a
solution in a 10 mm pathlength cuvette and get a reading of 2.0 AU, you can use a 5 mm
pathlength cuvette to get an absorbance of 1.0 AU for the same solution. SmartSpec Plus
corrects for pathlengths shorter than 10 mm when converting absorbances into concentration.

The following table gives you the absorbance at A260 and corresponding concentration

for the different pathlength cuvettes.

Pathlength

A260 = 0.1 AU

A260 = 1.5 AU

**

10 mm

5 µg/ml

*

75 µg/ml

5 mm

10 µg/ml

150 µg/ml

2 mm

25 µg/ml

375 µg/ml

1 mm

50 µg/ml

750 µg/ml

*

DNA concentration based on 1 O.D. = 50 µg/ml for dsDNA

**

Working range of the SmartSpec Plus is 0.1–1.5 AU per Good Laboratory Practice.

Cuvettes that can fit the SmartSpec Plus should have 8.5 mm 'Z' dimension and either 1,

2, 5 or 10 mm pathlength and 12.5 mm width. In general there are two types of 'Z' dimensions:
15 mm and 8.5 mm. This is only applicable for sub microvolume cuvettes i.e., for volumes
100 µl or lower. In most cases you will use 10 mm pathlength cuvettes.

The SmartSpec Plus has 8.5 mm 'Z' dimension. This is 8.5 mm from the bottom of the

cuvette and it is the distance required for the light path to hit the sample. See Figure below.