Recent Alarms page

Alarm History page

Table 13. WACS AcuityLink option: Clinician Notifier Recent Alarms page

Feature

Specification

Recent alarms shown

Active and inactive alarms, including arrhythmia events that
have been deleted from Acuity System review windows and
printouts

Situations when only a patient’s first alarm is shown

When a lower priority alarm occurs while another alarm
is asserting or suspended

When a higher priority alarm occurs while another alarm
is suspended, but the new alarm is neither a lethal
arrhythmia nor apnea alarm.

Number of alarms per patient stored in the page

Default: 25 alarms
Configurable 25-200 alarms

Recent alarms retention period

Default 4 hours
Configurable: 5 minutes-12 hours

Table 14. WACS AcuityLink option: Clinician Notifier Alarm History page

Feature

Specification

Alarms shown in Alarm History page

All of a patient’s most recent Acuity System alarms, including
arrhythmia events that have been deleted from Acuity System
review windows and printouts.

These types of alarms can be listed in the Alarm History page
without ever having appeared on the mobile device as an
alarm:

Alarm types not configured in the WACS Clinician Notifier
Configure Notification page to be delivered to mobile
devices

Low-level arrhythmia alarms (as set at Acuity System) are
not delivered to mobile devices running Clinician Notifier.

Short-duration alarms (shorter than the alarm delivery
holdoff variable in WACS Configure Notification page) are
not delivered to mobile devices running Clinician Notifier

Higher priority alarms that occur when another alarm is
suspended, but the new alarm does not interrupt the
suspend period because it is neither a lethal arrhythmia
nor an apnea alarm.

Lower priority alarms that occur while another alarm is
asserting or suspended

Multiple listings for the same alarm if the alarm’s popup
is replaced by another alarm that later ceases. The
original alarm is still active after the popup ceases.

Number of alarms per patient stored in the page

Default: 25 alarms
Configurable 25-200 alarms

Length of time alarm histories are shown after
disconnection from Acuity System

Default 96 hours
Configurable: 1-96 hours, but no longer than the Acuity
System Full Disclosure license period