3 hazardous location use (visolid® 700 iq h), Hazardous location use (visolid, Danger – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 11

Advertising
background image

ViSolid

®

700 IQ (SW)

Safety

ba76040e04

05/2014

2 - 3

2.3

Hazardous location use (ViSolid

®

700 IQ H)

Rated sensor models

The ViSolid

®

700 IQ H (YSI-Order No. 207012Y) is rated for hazardous

location use. Hazardous location rated sensors are identified by the
name plate with the rating details engraved on the sensor shaft:

Name plate of a Hazloc

sensor (example)

DANGER

Explosion hazard.
Only hazardous location rated sensor models
("Hazloc Sensor") must be used in hazardous
locations. Read the name plate on the sensor shaft
and verify that the sensor is rated for your specific
application.

DANGER

Explosion hazard.
The associated apparatus MIQ/BB2 (YSI part no.
207002Y) has to be used for connection. Refer to the
following documents for proper installation:
Control drawing, YSI document no. 28124

(see MIQ/BB2 operating manual)

MIQ/BB2 operating manual
All applicable electrical code regulations.

DANGER

Explosion hazard.
This product does not meet the requirements of the
directive 94/9/EC (ATEX).

######

######

######

ViSolid 700 IQ H

28124

#########
###############

####################
#################

Advertising