3 correction factor, Correction factor, See section 4.2.3) – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 33

Advertising
background image

ViSolid

®

700 IQ (SW)

Measuring

ba76040e04

05/2014

4 - 5

4.2.3

Correction factor

The setting of the Correction factor provides a simple option for
adapting the calibration to the current conditions.

With the Correction factor setting you correct the measured value and
have it indicated on the display.

A change of the Correction factor setting is practical if the measured
values of the ViSolid

®

700 IQ (SW) are generally too high or too low in

comparison to reference measurements by a specific factor.

Fig. 4-3

Effect of the Correction factor on the displayed measured value

The Correction factor is calculated using the following formula:

Variable

Explanations

F

N

Correction factor, to be recalculated

F

A

Correction factor, currently set
in the Calibration data menu

S

R

TSS value, newly determined from reference mea-
surement

S

V

TSS value, newly determined from measurement with
ViSolid

®

700 IQ (SW)

TSS [1000 mg/l]

SiO2 [1000 mg/l ]

0,9

0,7

0,5

0,3

0,1

Correction factor 1.00

Correction factor 0.50

F

N

= F

A

* S

R

/S

V