ViSolid
®
700 IQ (SW)
Commissioning
ba76040e04
05/2014
3 - 3
3.3.3
Flow direction
Generally, in flowing media, the measurement window should be
clearly pitched towards the flow (angle of incidence approx. 20 to 45 °).
Exception: If there is a high proportion of foreign bodies with fibrous or
flat profiles such as, e.g. hair, twines or foliage, it can be advantageous
to tilt the sensor in the direction of flow so that the measurement
window is turned away from the flow.
3.3.4
Sensor angle
Fig. 3-1
Effect of the sensor angle on scattering and reflection from the ground and
wall
Scattering and reflection are lowest at a sensor angle of
45° and at a minimum distance of 10 cm to the ground and
walls (see section 3.3.6).
0 °
20 °
45 °
10 cm
Side
view:
Marking