Commissioning

ViSolid

®

700 IQ (SW)

3.4.2

Measurement in pipelines

If deposits occur on the pipe walls, the calibration should be repeated
at regular intervals.

Example:

45 ° pipe installation

The pipe should be straight for a length of approx. 25 cm beyond the
installation location. Angled or tapered pipes can cause interference
effects in the case of low levels of total suspended solids.

Fig. 3-5

Total suspended solids sensor in the pipe with EBST 700-DU/N flow-thru
adapter

Fig. 3-5 shows the installation of the EBST 700-DU/N flow-thru adapter
for installation in a pipeline (DN 50). The infrared beam points in the
opposite direction to the direction of flow. The marking on the sensor
points towards the pipeline (see Fig. 3-5).

If there is a low level of total suspended solids (< 2000 mg/
l SiO2 or < 1000 mg/l TSS), the effects of the measurement
environment can simulate a higher content of total sus-
pended solids. The effect of the measurement environment
can be reduced by ensuring the optimum conditions (see
section 3.3.2).

If an optimum installation is not possible due to the struc-
tural conditions at the measuring location (e.g. in narrow
pipelines), the effects of the measurement environment
can be compensated by user calibration (see section
4.2.4)
.

For exceptions to the direction of flow, see section
3.3.3 F

LOW

DIRECTION

.

ADA-DF 9

EBST 700-DU/N

ca. 25 cm

straight pipe

Marking aid
in this direction

Swivel nut

Infrared beam

Direction

of flow