6 what to do if, What to do if -1, 6what to do if – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 43

Advertising
background image

ViSolid

®

700 IQ (SW)

What to do if ...

ba76040e04

05/2014

6 - 1

6

What to do if ...

Mechanical damage to

the sensor

Display always shows

"0"

TSS display does not

correspond to the TSS

value according to the

laboratory

determination

Display of OFL

Display of "----"

Display of main

parameter
TSS: "----"

Secondary parameter

SiO2: "OFL"

Measured value

fluctuating heavily

Cause

Remedy

– Return the sensor

Cause

Remedy

– First calibration value pair

incomplete

– Enter the TSS value for the

first calibration value pair

Cause

Remedy

– Correction factor incorrectly set

– Set up the correction factor

again: Correction factor =
TSS value (laboratory) /
TSS value (display)

Cause

Remedy

– Measuring range exceeded

– First calibration value pair

incomplete

– Two identical SiO2 values

entered one after the other

– See log book

– Enter the SiO2 value for the

first calibration value pair

– Enter the value pairs in

descending order

Cause

Remedy

– Measured value invalid

– Incorrect calibration value

entered

– See log book

– Correct the calibration val-

ues and enter them again

Cause

Remedy

– The optical measuring range for

SiO2 is exceeded. The display of
a valid TSS measured value is
not possible.

– See log book

(message code EA6243,
see section 8.1.1)

Cause

Remedy

– There are bubbles of gas in the

medium in front of the measure-
ment windows

– Check the installation posi-

tion of the sensor (see sec-
tion 3.3 and
section 3.4)