4 user calibration, User calibration, See section 4.2.4) – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 35: Section 4.2.4

Advertising
background image

ViSolid

®

700 IQ (SW)

Measuring

ba76040e04

05/2014

4 - 7

4.2.4

User calibration

The displayed values of total suspended solids are calculated with the
aid of the stored calibration data. In the mg/l TSS measuring mode, the
value mg/l SiO2 marked with "#" is displayed as the secondary mea-
sured value.
The determination of the calibration value pairs is carried out by refer-
ence measurements according to an independent procedure. At the
point of time of the calibration, the measuring medium should be in a
state representative of the later measurement (type and amount of
total suspended solids, coloration, etc.). The results from the calibration
are manually entered in the setting table of the ViSolid

®

700 IQ (SW).

Effect of the

measurement

environment on the

graph of the value pairs

TSS (laboratory) - SiO2

(ViSolid

®

700 IQ (SW)

)

Graphs of the value pairs

Explanations

Form 1

:

The graph has a gradient > 0 at every
point.
A calibration is possible for the whole
range.

Form 2

:

The graph has a gradient > 0 at every
point.
At very low levels of TSS, the effect of
the measurement environment leads
to slightly increased values of SiO2.
A calibration is possible for the whole
range.

Form 3

:

The graph has a gradient < 0 in the
range of smaller levels of TSS.
The effect of the measurement envi-
ronment leads to greatly increased val-
ues of SiO2 in the range of smaller
TSS below the point (A). A calibration
is only possible in the range TSS > A.

Form 4

:

The graph has a gradient < 0 in the
range of greater levels of total sus-
pended solids.
A calibration for matrix type 1 is only
possible in the range of TSS < B.
For measurements in the range of
TSS > B, select matrix type 2.

SiO2

[mg/l]

TSS [mg/l]

SiO2

[mg/l ]

TSS [mg/l]

SiO2

[mg/l ]

TSS [mg/l]

A

SiO2 [mg/l]

TSS [mg/l] B