YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 34

Advertising
background image

4 - 6

ba76040e04

05/2014

Measuring

ViSolid

®

700 IQ (SW)

Determining the

Correction factor

Setting the Correction

factor

1

Bring the sensor into the measuring position.

2

In the setting table of the TSS sensor, note down the currently
set Correction factor as the value for F

A

.

3

Switch to the measured value display with <M>.

4

When the measured value is stable, read the TSS value, con-
vert it into the unit (mg/l) if necessary,
and note it down as the value for Sv.

5

Take a sample as close to the same time as the TSS measure-
ment as possible and as close to the sensor as possible.

6

Determine the level of total suspended solids of the sample
according to a reference procedure (e. g. gravimetric accord-
ing DIN 38414), convert it into the unit (mg/l) if necessary,
and note it down as the value for S

R

.

7

Calculate the Correction factor.

F

N

= F

A

* S

R

/S

V

8

Set the new Correction factor in the Calibration data menu (see
section 3.5.3).

The Correction factor setting affects each TSS measuring
mode and all calibration data. Also, if there is a change of
the measuring mode or calibration data, the Correction fac-
tor

is retained. Therefore, after every change of the set-

tings in the Calibration data menu, check the Correction
factor

.

Advertising