700 IQ (SW)

Technical data

ba76040e04

05/2014

7 - 1

7

Technical data

7.1

Measuring characteristics

Measuring principle

Procedure for measuring scattered light.
Measurement in following units:

mg/l TSS (total suspended solids)

% TSS (total suspended solids)

mg/l SiO

2

% SiO

2

Measuring ranges and

resolutions

Measured parameter

Measuring ranges

Resolution

mg/l TSS

0 ... 400.0 mg/l
0 ... 4000 mg/l
0 ... 25000 mg/l
0 ... 40000 mg/l

0.1 mg/l
1 mg/l
1 mg/l
1 mg/l

% TSS

0 ... 400.0 ppm
0 ... 4000 ppm
0 ... 2.500 %
0 ... 4.000 %
0 ... 40.00 %
0 ... 100.0 %

0.1 ppm
1 ppm
0.001 %
0.001 %
0.01%
0.1%

mg/l SiO

2

0 ... 400.0 mg/l
0 ... 4000 mg/l
0 ... 25000 mg/l
0 ... 40000 mg/l

0.1 mg/l
1 mg/l
1 mg/l
1 mg/l

% SiO

2

0 ... 400.0 ppm
0 ... 4000 ppm
0 ... 2.500 %
0 ... 4.000 %
0 ... 30.00 %

0.1 ppm
1 ppm
0.001 %
0.001 %
0.01 %