10 appendix, 1 check calibration values, 10 appendix -1 – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 57: 1 check calibration values -1

Advertising
background image

ViSolid

®

700 IQ (SW)

Appendix

ba76040e04

05/2014

10 - 1

10

Appendix

10.1

Check calibration values

By checking the value pairs, possible calibration errors can already be
avoided before the entry of the calibration value pairs.

Carry out a check using the EC2342 message text:

* All TSS values within measuring range?
(see operating manual)
* At least one value pair entered?
* All TSS and SiO2 values entered?
* All TSS/SiO2 pairs in descending order?
* Pair 1 = highest TSS and SiO2 value?

Sequence

Max. value

->

Min. values

Value pairs

1

2

3

4

5

6

7

8

Total suspended solids
[mg/l] TSS

SiO2 value [mg/l] SiO2

For examples of valid value pair data, see section 4.2.2.
The graph of the value pairs must not contain any turning
point in the required measuring range (see section 4.2.4).

TS [1000 mg/l]

SiO2 [1000 mg/l]

10

0,9

8

0,7

6

0,5

4

0,3

2

0,1

0

0

2

4

6

8

10

12

14

16

18

20