Commissioning

ViSolid

®

700 IQ (SW)

3.5.2

Selecting the Measuring mode

Specify the following data in the Measuring mode setting

Matrix type (1 or 2)

Display (TSS or SiO2)

Unit (mg/l or %)

Determining the matrix

type

Determine the matrix type for your application with the aid of the follow-
ing table:

Measurement in mg/l

TSS (total suspended

solids)

Measurement in mg/l

SiO2 (silicon dioxide)

Factory settings

Measuring mode: Mtrx.type1:mg/l TSS
Measuring range: AutoRange

Measuring medium:

Matrix type

1

Matrix type

2

Outflow of preclarification

X

Activated slurry

X

Return slurry

X

Primary slurry

X

Sediment slurry

X

Concentrated slurry

X

Measuring medium:

Matrix type

1

Matrix type

2

SiO2 content: 0 ... 25000 mg/l

X

SiO2 content: 15000 ... 300000 mg/l

X

If the matrix type for a measuring medium cannot be deter-
mined using this table, select matrix type 1 and then carry
out a user calibration (see section 4.2.4). If the graph of the
value pairs corresponds to one of the forms 1 to 3 (see sec-
tion 4.2.4), matrix type
1 is suitable. If the graph of the value
pairs corresponds to form 4, matrix type 2 must be
selected.