2 application characteristics, Application characteristics -2 – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual
Page 46
7 - 2
ba76040e04
05/2014
Technical data
ViSolid
®
700 IQ (SW)
7.2
Application characteristics
Allowed
temperature range
Allowed pH range of the
measuring medium
4 ... 12
Pressure resistance
Sensor with connected SACIQ (SW) sensor connection cable:
The sensor fulfills all requirements according to article 3(3) of the direc-
tive 97/23/EC ("pressure equipment directive").
Type of protection
Sensor with connected SACIQ (SW) sensor connection cable:
IP 68, 10 bar (10
6
Pa)
Depth of immersion
min. 10 cm; max. 100 m depth
Operating position
Fields of application
Water and wastewater monitoring
Measuring medium
0 °C ... + 60 °C (32 ... 140 °F)
Storage/transport
- 5 °C ... + 65 °C (23 ... 149 °F)
Max. allowed overpressure
10
6
Pa (10 bar)
Max. allowed underpressure
temporarily 5 x 10
4
Pa (0,5 bar)