YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 21

Advertising
background image

ViSolid

®

700 IQ (SW)

Commissioning

ba76040e04

05/2014

3 - 9

Example:

90 ° pipe installation

Fig. 3-6

Total suspended solids sensor in a pipe (90 °)

The following points must be observed for a right-angled installation in
the pipe (Fig. 3-6):

Rotate the sensor so that the marking on the sensor points in the
direction of the pipe axis

Select a position where the pipe diameter is as large as possible as
the installation location (see section 3.3.6 D

ISTANCES

FROM

THE

GROUND

AND

WALL

).

Marking aid

In a 90° pipe installation with low levels of total suspended
solids (< 2000 mg/l SiO2 or < 1000 mg/l TSS), the effects
of the measurement environment can have a particularly
significant effect on the measured value. Ensure the opti-
mum conditions of the measurement environment (see
section 3.3.2).

Infrared beam

EBS 700-DU/N

Marking aid
in this direction

ADA-DF 9

1

Connect the SACIQ (SW) sensor connection cable to the plug
head connector of the sensor and screw it tight (see section
3.5.1)
.

2

Attach a marking aid (adhesive strips or similar) in the same
position as the marking on the sensor to the plug head connec-
tor.