YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual
Page 44
Advertising
6 - 2
ba76040e04
05/2014
What to do if ...
ViSolid
®
700 IQ (SW)
Measured values too
low
Measured values too
high
– Signal averaging time too short
for low values of total suspended
solids
– Inhomogeneous measuring
medium
– Increase signal averaging
time
Cause
Remedy
Cause
Remedy
– Measurement window dirty
– Clean the measurement
Cause
Remedy
– There are bubbles of gas in the
medium in front of the measure-
ment windows
– Check the installation posi-
tion of the sensor (see sec-
tion 3.3 and section 3.4)
– Light scattering on the walls
– Check the installation posi-
tion of the sensor (see sec-
tion 3.3 and section 3.4)
– If necessary, compensate
for any effects that cannot
be cleared by calibration
– Measurement window dirty
– Clean the measurement