YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 26

Advertising
background image

3 - 14

ba76040e04

05/2014

Commissioning

ViSolid

®

700 IQ (SW)

TSS measuring
mode:
Menu selection:
User calibration

Basic settings

Value pairs 1..3
Value pairs 4..6
Value pairs 7..8

Selection between the use of the basic set-
tings and the entry of calibration value pairs.

For the selection of the value pairs, fields
open for the entry of the values
TSS value 1

to TSS value 8.

and the individual values
SiO2 value 1

to SiO2 value 8.

Starting with TSS value 1, determine and
enter the content of total suspended solids in
mg/l TSS, and, in the case of SiO2 value 1,
enter the related measured SiO2 value.

The input precision is 1 mg/l in each case.

Note:
The values must be entered in descending
order.
If this sequence is not kept to, a calibration
error is displayed after leaving the menu. All
the entered values become invalid.

TSS measuring
mode:
Measuring ranges

The setting
AutoRange =
automatic change-
over of the measur-
ing range
can be selected in all
measuring modes.

0 ... 400.0 mg/l
0 ... 4000 mg/l
0 ... 25000 mg/l

Measuring range for the measuring mode
Mtrx.type1:mg/l TSS

0 ... 400.0 ppm
0 ... 4000 ppm
0 ... 2.500 %

Measuring ranges for the
Matrix type1:% TSS

measuring mode

0 ... 4000 mg/l
0 ... 40000 mg/l

Measuring ranges for the
Mtrx.type2:mg/l TSS

measuring mode

0 ... 4000 ppm
0 ... 4.000 %
0 ... 40.00 %
0 ... 100.0 %

Measuring ranges for the
Matrix type2:% TSS

measuring mode

Setting

Selection/values

Explanation