Measuring

ViSolid

®

700 IQ (SW)

The calibration data entered are evaluated by the system. The entry of
the calibration value pairs can have the following results:

Results after entry of

Calibration data

17

Enter the value for the contents of total suspended solids (TSS
in mg/l) from the reference measurement with <> and
<OK>

.

18

Select the SiO2 value 1 menu item with <> and <OK>.

19

Enter the associated SiO2 value (SiO2 in mg/l) measured with
the ViSolid

®

700 IQ (SW) with <> and <OK>.

20

Repeat steps 13 - 18 until the required number of value pairs
(between 1 and 8) has been entered.

21

Terminate the entry of the calibration data with Save and quit.

To simplify the settings when entering the calibration data,
standardized high resolutions of 1 mg/l are possible. How-
ever, it is not sensible to use these in all cases.

Possible displays

Log book entries
(meaning/actions)

Measured value display

Successful new valid calibration value pairs
were entered for the sensor.

"----"

Calibration value pairs not accepted.
Sensor blocked for measurement.

– Enter the values again making sure that

they are entered in descending order

– Viewing the log book entry.

Information on the contents and structure of the log book
and how you can call it up is given in the L

OG

BOOK

chapter

of the IQ

S

ENSOR

N

ET

system operating manual.