8 indexes, 1 explanation of the messages, 1 error messages – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 51: Indexes -1, Explanation of the messages -1 8.1.1, Error messages -1, 8indexes

Advertising
background image

ViSolid

®

700 IQ (SW)

Indexes

ba76040e04

05/2014

8 - 1

8

Indexes

8.1

Explanation of the messages

This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ

S

ENSOR

N

ET

system for the ViSolid

®

700 IQ (SW) sensor.

8.1.1

Error messages

Information on

the contents and structure of the log book and

the structure of the message code

is given in the

L

OG

BOOK

chapter of the IQ

S

ENSOR

N

ET

sys-

tem operating manual.

All message codes of the ViSolid

®

700 IQ (SW) end with

the number "342".

Message code

Message text

EA2342

Sensor temperature too high!
* Check process and application

EA3342

Sensor temperature too low!
* Check process and application

EA6342

Meas. range exceeded or undercut
* Check process
* Select other meas. range
* Submerse sensor in sample
* Select bubble-free spot for measurement
* Remove any foreign matter from sensor
* Avoid influence of large foreign matter
* Clean sensor
* Increase signal average time

EA7342

Ultrasound cleaning system switched off
* Check sample temperature
* Submerse sensor in sample

Advertising