2 default calibration, See section 4.2.2) – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 31

Advertising
background image

ViSolid

®

700 IQ (SW)

Measuring

ba76040e04

05/2014

4 - 3

4.2.2

Default calibration

Default calibration for

matrix type 1

The factory calibration curve for matrix type 1 was determined by
measurements of typical activated and return slurries and can be used
for similar applications after adaptation of the Correction factor setting
(see section 4.2.3).

Fig. 4-1

Default calibration for matrix type 1

Below the smallest value, the calibration curves are
extended to the zero point and, above the largest value,
they are extended to the end of the measuring range.

Value pairs

1

2

3

4

5

6

7

8

9

10

Total suspended solids
[1000 mg/l] TSS

17.57 15.55 11.62 8.80 6.21 4.42 3.39 2.40 0.77 0.25

SiO2 value
[1000 mg/l] SiO2

7.16

7.05

6.52 5.85 4.86 3.91 3.22 2.60 1.37 0.61

TS [1000 mg/l]

SiO2 [1000 mg/l ]

25

0,9

20

0,7

15

0,5

10

0,3

5

0,1

0

0

1

2

3

4

5

6

7

8

9

10