3 setting tablevisolid® 700 iq (sw), Setting tablevisolid – YSI IQ SensorNet ViSolid Sensor 700 IQ User Manual

Page 25

Advertising
background image

ViSolid

®

700 IQ (SW)

Commissioning

ba76040e04

05/2014

3 - 13

3.5.3

Setting table

ViSolid

®

700 IQ (SW)

Setting

Selection/values

Explanation

Measuring mode
(see section 3.5.2)

Mtrx.type1:mg/l TSS
Matrix type1:% TSS
Mtrx.type2:mg/l TSS
Matrix type2:% TSS
Mtrx.type1:mg/l SiO2
Matrix type1:% SiO2
Mtrx.type2:mg/l SiO2
Matrix type2:% SiO2

– Content of total suspended solids in mg/l

– Content of total suspended solids in %

– Content of total suspended solids in mg/l

– Content of total suspended solids in %

– Content of SiO

2

in mg/l

– Content of SiO

2

in %

– Content of SiO

2

in mg/l

– Content of SiO

2

in %

Signal averaging

1 ... 600 secs

Response time of the signal filter.
Depending on the sample matrix, the mea-
sured values may oscillate more or less (e.g.
due to foreign bodies or air bubbles).
The signal filter reduces the limits of variation
of the measured value. The signal filter is
characterized by the signal averaging time.
This is the time after which 90 % of a signal
change is displayed.

Ultrasonic cleaning

On / On
Off / On
Pulse / On

On / Off
Off / Off

Pulse / Off

Switches on or off the ultrasound cleaning
and SenseCheck functions (Pulse = pulse
operation).

Save and quit

The system confirms the saving of the set-
tings and the display switches to the next
higher level.

Quit

The display switches to the next higher level
without saving the new settings.

TSS measuring
mode:
Calibration data

Default calibration

The factory calibration data for TSS are used
(see section 4.2.2).

User calibration

User entered calibration data are used (see
section 4.2.4).

TSS measuring
mode:
Correction factor

0.50 ... 2.00

The Correction factor setting enables a sim-
ple calibration (see section 4.2.3).