Maintenance, cleaning, accessories

ba76040e04

05/2014

5 - 1

5

Maintenance, cleaning, accessories

5.1

General information

The ViSolid

®

700 IQ (SW) sensor does not usually require any mainte-

nance. The continuously running ultrasound system prevents the accu-
mulation of pollution right from the start.

5.2

Cleaning the sensor shaft and measurement windows

During normal operation (e.g. municipal wastewater), cleaning is rec-
ommended:

if there is any pollution (according to visual check)

if the sensor was not in operation for a lengthy period of time but was
immersed in the measuring medium

if the measured values are suspected of being incorrect (usually too
low)

if the SensCheck message appears in the log book (when using
samples of matrix type 1)

for regular cleaning (when using samples of matrix type 2)

if there is any suspicion that the measurement window is polluted,
e.g. by dried on dirt during operation in the open air

Cleaning agents

WARNING

Chemical or biological hazard.
Contact with the sample can be harmful to the user.
Depending on the type of sample, suitable protective
measures must be taken (protective clothing, gog-
gles, etc.).

If the sensor remains in the sample for any length of time
when the system is not operating, we recommend to clean
the shaft and measurement windows.

Contamination

Cleaning agents

Sludge and loosely adhering dirt
or biological films

Soft cloth or soft brush, warm tap
water with detergent

Salt and / or lime deposits

Acetic acid (volume percentage
= 20 %), soft cloth or soft sponge