YSI 6200 User Manual

Page 106

Advertising
background image

YSI 6200 DAS

USER Manual

YSI/Massachusetts 508.748.0366, Fax 508.748.2543

Page 9-14

Phone modem Communication

The phone modem is set to factory default profile.

SYMPTOMS

POSSIBLE CAUSE

ACTION

REF

Depleted battery

Check battery voltage, replace
battery if necessary.

4.3

Damaged modem at base
station or corrupted modem
profile

Check with another modem, use
terminal emulator to check modem
profile

4.4

Phone modem

6200 DCP will not
respond to the base
station modem

Damaged modem at site or
corrupted modem profile

Visit site, verify 6200 DCP is
operational, and observe that
modem LED turns on during an
incoming call. If LED never turns
on, then modem is inoperative or
phone line is damaged. Test line
with another phone.
If the phone line is OK, replace the
modem.

4.4

Modem connects,
no data or 6200
DCP user screen
displayed

6200 DCP modem port
control pins not detecting
the ring, thus, the 6200DCP
can not strobe the power to
the modem.

6200 DCP interface board
damaged, return to factory

4.4