YSI 6200 DAS USER Manual

YSI/Massachusetts 508.748.0366, Fax 508.748.2543 Page 5-9

6504 Breakout Box

The 6504 Breakout Box is designed to convert one cable with SDI-12 information to two MS-8
Sonde connectors. This box works well with the 6507 Patch Cable. You will need a waterproof
gland to fit on this cable. Feed the patch cable through this gland fitting and wire into the TB-1
terminal.

Wiring Diagram

(1) Red

+12 VDC

(2) Black

Common

(3) Purple or Blue

SDI-12

Figure 5.7 Standard Junction Box Connection, Pin Outs and Wire Colors

Junction Boxes

Junction Boxes are used whenever a cable connection is needed. The 6508 Junction Box has a
terminal strip inside to help facilitate these connections, or they can be done with wire nuts.
Always use some desiccant to be sure the connections remain dry.

The connection, including pin outs and wire colors are shown in Figure 5.8.

Wiring Diagram

(A) Red

+12 VDC

(B) Black

Common

(F) Purple or Blue

SDI-12

Figure 5.8 Standard Junction Box Connection, Pin Outs and Wire Colors

(2) MS-8 to Sonde

3/4” conduit to Jct. Box

MS-8

Sonde Cables “in”

Junction Box

+12 vdc

SDI-12

Gnd