YSI 6200 User Manual

Page 98

Advertising
background image

YSI 6200 DAS

USER Manual

YSI/Massachusetts 508.748.0366, Fax 508.748.2543

Page 9-6

6200 DCP

The first method to recover a lost 6200 DCP system is to use the AutoConfig function.
This will result in the loss of data that has already been collected; however, it will check
all sensors, re-program, and reset the system.

The 6200 DCP contains the circuit board that collects all of the data and the
communication methods. This circuit has an EEPROM that is programmed by
EcoWatch DCP to carry out it’s tasks. Generally, every time a change is made with
EcoWatch, the 6200 DCP is re-programmed completely and re-booted. The COM port
located inside the enclosure will take priority over any other communication or task that
the 6200 DCP is doing. To check out the field station, load a laptop with EcoWatch and
transfer the configuration files from the base station to the laptop. This is done by
backing up the configuration to disk, then restoring that configuration from disk to the
laptop. You will then have a configuration that is ready to talk to the field station.

At the field station, connect with a straight thru RS-232 cable (provided). In EcoWatch,
switch to Direct, and use the Interrogate Now function. You should hear a small click
from the 6200 DCP as it switches over to communicate thru the front panel COM port.

SYMPTOMS

POSSIBLE CAUSE

ACTION

REF

Depleted battery

Check battery voltage, replace
battery if necessary. 11.7 VDC is
needed for boot-up.

4.3

Program is looping

Remove power for 5 seconds, then
re-apply. You should hear a beep
for about 5 seconds at boot-up.

4.3

Wrong COM port on laptop

Change as necessary.

2

Corrupted configuration file

AutoConfig to reset the program in
the 6200 DCP (6200:AutoConfig).
If this does not work, begin a New
file.

7.6

6200 DCP does not
respond

Check communication
method

Refer to communication section
below.

6.6

Program has changed

AutoConfig to reset program in 6200
DCP (6200:AutoConfig)

7.6

Check sensors

A sensor may be disconnected or
fouled and needs maintenance.

5

RF influencing Sondes

Be sure Sonde cables are not run
near other power lines. Shield if
necessary

7.2

6200 DCP is giving
bad data

Conflicting configurations

If more than one base station
communicates with 6200 DCP, they
must all have the same config files.

6.6

6200 DCP
Continuously beeps

Program is lost

Connect to computer with EcoWatch
DCP and use 6200:Advanced:
Program Now function.
AutoConfig after to be sure program
is correct.

7.6